Home Frequently Asked Questions Tracker 5.0 How do I full calibrate my instruments?
How do I full calibrate my instruments? Print E-mail
User Rating: / 0
PoorBest 
Thursday, July 01 2010 06:27

A full calibration is recommended for all JTECH force gauges on a yearly basis with certified weights. For information on JTECH factory calibration, contact JTECH Customer Service.

1. Select Full Calibration from the system drop-down menu and
the instrument for calibration.

2. Wake up a wireless instrument by gently pressing the “On” button. For Legacy instrument calibration, make sure the Tracker Box is turned on.

Note: The value displayed in the upper right-hand corner of the Calibration screen is the raw value that will be captured when the calibration is performed. A Ø symbol indicates that communication with the instrument has not been established.

3. Make sure there is no weight or force applied to the instrument and click the Zero button to zero calibrate the device.

4. Enter the amount of weight for the calibration into the Weight field and select the “Units” (pounds, metric or newtons) from
the drop-down list. Normal ranges for weight calibrating the device are displayed in the screen’s right-hand column.

5. Apply the weight to the instrument and click the Calibrate button.

4. Compare the current calibration data with the previous calibration data and the factory calibration data. If the data varies more than 5%, contact JTECH Customer Service.